If you are wondering which languages are required for your company to place medical devices on the market in the European Union, the answer depends on the countries where you plan to sell them, the type of information involved and the intended end user of the medical device.
It is also important to know that there are exemptions. Some countries accept English for certain documents or medical devices intended for healthcare professionals.
Software interfaces require particular attention because the MDR does not contain a specific article addressing their language requirements, and some countries do not require them to be translated.
Therefore, to define the scope of translation, you need to identify the requirements that apply in each target market. The European Commission’s MDR and IVDR language requirements tables provide an overview of the differences between countries, documents and user types, including specific information on user interfaces. They are not entirely up to date, but they provide a good starting point.
Do you need to translate into all official EU languages?
The EU has 24 official languages: Bulgarian, Croatian, Czech, Danish, Dutch, English, Estonian, Finnish, French, German, Greek, Hungarian, Irish, Italian, Latvian, Lithuanian, Maltese, Polish, Portuguese, Romanian, Slovak, Slovenian, Spanish and Swedish. This list describes the EU’s language regime, but does not in itself establish which translations are mandatory for your devices.
As we will see in the next section, the scope depends on where you market them. For this reason, budgeting for ‘all official EU languages’ without first reviewing the documentation can lead to unnecessary costs.
How can you determine which languages your medical device documentation needs to be translated into?
Article 10(11) of Regulation (EU) 2017/745 on medical devices, known as the MDR, stipulates that labelling and instructions for use, as defined in Section 23 of Annex I, must be provided in the official EU languages determined by the Member State in which the device is made available to the user or patient.
The starting point is therefore to identify markets where you plan to sell your devices. Next, you need to check the requirements in each country for labelling, instructions for use and any other applicable information. It is also important to distinguish between medical devices intended for healthcare professionals and those intended for patients or people without healthcare training (referred to as lay persons in the MDR).
For in vitrodiagnostic devices, refer specifically to the corresponding IVDR table. While the approach is similar, it is advisable to check each regulatory framework separately. The European Commission’s language requirements tables serve as a useful reference tool. However, bear in mind that the applicable legislation takes precedence over any other reference source, including MDCG guidance.
The following table sets out the languages into which you will need to translate labelling and instructions for use, regardless of the risk class of your medical devices:
| Country | Languages | Exemptions | National Legislation |
|---|---|---|---|
| Germany | German | For devices intended solely for professional use, English or another language understood by the user may be accepted if justified; safety information must also be provided in German or in the user’s language. | Medizinprodukterecht-Durchführungsgesetz (MPDG) |
| Austria | German | For devices intended exclusively for healthcare professionals, the information may be provided in German or English. | Medizinproduktegesetz 2021 (MPG 2021) |
| Belgium | French, Dutch and German | English is also accepted for professional users; for lay users, the applicable national languages are required. Professional users may request the information in the national language of their choice. | Wet van 22 december 2020 betreffende medische hulpmiddelen, art. 9 |
| Bulgaria | Bulgarian | No general exemptions. | Закон за медицинските изделия |
| Czechia | Czech | No general exemptions. | Zákon č. 375/2022 Sb. o zdravotnických prostředcích a diagnostických zdravotnických prostředcích in vitro |
| Cyprus | Greek | Greek or English is accepted for professional users. The national legislation dates back to the Directives and its application to the MDR has not yet been confirmed. A new regulation was submitted for consultation in 2024. | Ν. 30(I)/2002 και Κ.Δ.Π. 598/2003 |
| Croatia | Croatian | For professional use, Croatian and/or English may be used, subject to the applicable conditions and agreements. | Zakon o provedbi uredbi (EU) 2017/745 i 2017/746 |
| Denmark | Danish | The Danish Medicines Agency may exceptionally authorise another language on public health grounds. This authorisation is not available for self-testing or near-patient in vitro diagnostic devices. | Bekendtgørelse nr. 1731 af 8. december 2025 om medicinsk udstyr (replaces No. 837/2023) |
| Slovakia | Slovak | The labelling of devices intended for professional use may be in English, provided the devices are not intended for patients; instructions for use must always be in Slovak. | Zákon č. 362/2011 Z. z. o liekoch a zdravotníckych pomôckach, § 110b |
| Slovenia | Slovenian | For devices intended for professional use, a language understood by the user, usually English, may be used. If requested by the user, the instructions for use must be provided in Slovenian within 10 days. | Zakon o medicinskih pripomočkih (ZMedPri-1), art. 7 |
| Spain | Spanish | No general exemptions. | Real Decreto 192/2023, de 21 de marzo, por el que se regulan los productos sanitarios |
| Estonia | Estonian | English may be accepted for professional use, although the user may request an Estonian version. | Meditsiiniseadme seadus |
| Finland | Finnish and Swedish | English may also be accepted for professional users; information necessary for safe use must be provided in Finnish and Swedish, as determined by the manufacturer through a risk analysis. | Laki lääkinnällisistä laitteista 719/2021 |
| France | French | No general exemptions. | Code de la santé publique, art. R. 5211-3 (amended by Decree no. 2026-730) |
| Greece | Greek | For devices intended exclusively for healthcare professionals, labelling and instructions for use may be provided in English if authorised by the EOF upon application (Circular 37580/31.03.2025, published on 11 June 2025). This is not a general exemption from the requirement to translate into Greek. The MDR is the current regulatory framework; the national decisions from 2009 date back to the Directives era. | ΔΥ8δ/Γ.Π.οικ.130648/2009 (ΦΕΚ Β’ 2198), art. 4(4), and ΔΥ8δ/Γ.Π.οικ.130644/2009 (ΦΕΚ Β’ 2197); EOF Circular 37580/31.03.2025 |
| Hungary | Hungarian | No exemption for professional use. According to the competent authority, the software interface may be in English if the Hungarian instructions for use explain it through screenshots. | 4/2009. (III. 17 EüM rendelet az orvostechnikai eszközökről, § 5 |
| Ireland | English (mandatory); Irish (optional) | Irish may be used alongside English, but cannot replace it (S.I. 547/2017, art. 5(a)). | S.I. No. 547/2017 — European Union (Medical Devices and In Vitro Diagnostic Medical Devices) Regulations 2017 |
| Italy | Italian | No exemption applies to professional use: the information must be provided in Italian when the device is delivered to the end user. | Decreto legislativo 5 agosto 2022, n. 137, art. 6 |
| Latvia | Latvian | For devices intended exclusively for professional use, English may be accepted with the consent of the healthcare institution using the device. | Ministru kabineta noteikumi Nr. 461 «Medicīnisko ierīču noteikumi» |
| Lithuania | Lithuanian | Since 21 June 2024, the instructions for use of devices intended exclusively for healthcare professionals may be provided in English with the agreement of the health institution. | Lietuvos Respublikos sveikatos sistemos įstatymas Nr. I-552, 59⁵ straipsnis |
| Luxembourg | French, German or Luxembourgish | English may also be accepted for professional users; information intended for patients must be provided in a language they understand. For active implantable medical devices, only French or German is accepted. | Règlement grand-ducal du 11 août 1996 relatif aux dispositifs médicaux |
| Malta | Maltese and/or English | Labelling and instructions for use may be provided in Maltese and/or English. | S.L. 458.59 — Medical Devices and In-Vitro Diagnostic Medical Devices Provision on the Maltese Market Regulations |
| Netherlands | Dutch | Dutch or English is accepted for professional users, provided they have a sufficient command of English; Dutch is required for lay users. | Regeling medische hulpmiddelen |
| Poland | Polish | English may be used for healthcare professionals, whereas Polish is mandatory for lay users. | Ustawa z dnia 7 kwietnia 2022 r. o wyrobach medycznych |
| Portugal | Portuguese | No exemption applies to professional use; other languages may be provided in addition to Portuguese. | Decreto-Lei n.º 29/2024, de 5 de abril, art. 40.º |
| Romania | Romanian | For professional use, English may be accepted with the approval of the competent authority and the written consent of the healthcare professional, in accordance with the procedure in Ministry of Health Order No. 3539/2022. | Ordonanța de urgență nr. 46/2021 privind punerea în aplicare a Regulamentului (UE) 2017/745 |
| Sweden | Swedish | No exemption applies to professional use; the Swedish Medical Products Agency (Läkemedelsverket) may grant exemptions in specific cases and subject to certain conditions. | Förordning (2021:631) med kompletterande bestämmelser till EU:s förordningar om medicintekniska produkter |
| Iceland (EEA) | Icelandic | For Class I and IIa devices, information intended for lay users may be provided in English or another Nordic language (except Finnish). For professional use, Icelandic or English is accepted. | Lög nr. 132/2020 um lækningatæki, art. 12 |
| Liechtenstein (EEA) | German | English may be accepted if the device is supplied exclusively to healthcare professionals and all other legal requirements are met. | EWR-Medizinprodukteverordnung (EWR-MepV), LGBl. 2021 Nr. 161, art. 10 |
| Norway (EEA) | Norwegian | Norwegian is also required for professional use; the Norwegian Medical Products Agency (DMP) may grant exemptions if safe and proper use can be ensured. | Forskrift 9. mai 2021 nr. 1476 om medisinsk utstyr, § 6 |
| Turkey | Turkish | Labelling for devices intended for professional use may be provided in English with the approval of the competent authority (TİTCK), in accordance with Circular No. 2022/1. Instructions for use must be provided in Turkish. Turkey is not part of the EEA and applies its own national regulation (TR-MDR), which is harmonised with the MDR. | Tıbbi Cihaz Yönetmeliği (RG 2/6/2021, No. 31499), art. 10, and Law No. 7223, art. 7 |
Source: European Commission, language requirements for manufacturers under the MDR, revision 3 (August 2025), cross-checked against the national legislation in force as of October 2026. The Commission’s table does not contain up-to-date information for Austria, Bulgaria, Hungary, Italy or Lithuania, and the Danish order it cites has been superseded. Iceland, Liechtenstein and Norway apply the MDR as members of the European Economic Area (EEA), whereas Turkey has its own national regulation aligned with the MDR. Indicative summary: check the national legislation currently in force, the type of user and the specific document.
Which documents need to be translated?
Start by identifying the information supplied with the device and the information intended for users, whether healthcare professionals or lay persons. Then review the documents intended for competent authorities or notified bodies. Separating these two groups will help you define a more precise scope.
The main content to review includes:
| Document or content | What to check | Recipient |
|---|---|---|
| Labelling and instructions for use | Possible exemptions. | Intended user: healthcare professional, patient or lay person, depending on the device. |
| Implant card and patient information | Whether they apply and which languages are required in the target country. | Patient who has received the implant. |
| EU declaration of conformity | The languages accepted for this specific document. | Competent authorities; authorised representative, importers and distributors, according to their respective obligations. |
| Field safety notices | The requirements applicable in the markets covered by the communication. | Intended users: healthcare professionals and institutions and, where applicable, patients. |
| Summary of safety and clinical performance (SSCP/SSP) | Whether it is applicable and which language versions are required. | Intended users and, where applicable, patients; publicly available. |
| Conformity assessment documentation | The language requirements applicable to the procedure and the intended recipient. | Notified body, where involved, and the competent authorities requesting the documentation. |
The MDR establishes distinct requirements for these documents. There is no general obligation to translate the full technical documentation into all the languages of the countries where the device will be placed on the market. Before commissioning any translations, it is worth confirming which documents need to be translated.
For the summary of safety and clinical performance, MDCG 2019-9 guidance recommends translating it into the languages accepted in the Member States where the device is sold, with a separate document for each language.
The notified body validates only one language version. If that version is not in English, the guidance states that the manufacturer should provide an English translation within 90 days. The remaining translations are not reviewed by the notified body, so their accuracy depends on the manufacturer's quality management system.
For the implant card, MDCG 2019-8 guidance recommends using symbols to avoid the need for national language versions, although the product type field must be provided in the language of each country.
When preparing the translation of the instructions for use (IFU), it is advisable to also provide any related labels, illustrations and on-screen text, where applicable. This makes it easier to maintain consistent terminology for components, warnings and user actions.
Examples of cases where translation into the national language is not required
Some countries accept alternatives for certain documents or devices intended for professional use. These options must be checked against the specific conditions that apply. The fact that a hospital routinely operates in English is not, by itself, sufficient grounds for omitting a translation.
The following examples, taken from the European Commission’s table of language requirements for manufacturers under the MDR, revision 3, August 2025, illustrate why each combination of country, document and intended recipient needs to be assessed individually:
- Netherlands: the MDR table distinguishes between information intended for lay persons, which must be in Dutch, and information intended for professional users, which may be provided in Dutch or English.
- Malta: the table lists Maltese and/or English for labelling and instructions for use.
- Germany: there are specific conditions under which certain information intended exclusively for healthcare professionals may be provided in English or another language they understand.
In Germany, Article 8 of the national medical devices legislation requires these cases to be justified. In addition, safety information must also be available in German or in the user’s language. The same article allows the EU declaration of conformity to be drawn up in German or English, demonstrating that language requirements may differ from one document to another.
Before applying an exemption, document the applicable legal basis and the conditions your device meets. Do not apply a language exemption to other documents, users or markets without checking that it is also valid in those cases.
Translation of medical software interfaces (SaMD): a separate check
Interfaces warrant a separate review, both in software as a medical device (SaMD) and in software that operates as an accessory to medical equipment. This is because translation can be costly and the regulations provide little guidance on whether these interfaces need to be translated.
In the ‘Graphic User Interface (GUI)’ section of the page Overview of language requirements for manufacturers of medical devices, the European Commission clarifies that the MDR and IVDR do not contain a specific article dedicated to language requirements for user interfaces.
The same page explains that some countries have national legislation and rules governing these interfaces, with the relevant requirements listed in the last column of the tables available for both the MDR and the IVDR. Therefore, translating the instructions for use does not automatically ensure compliance with the interface language requirements.
In the ‘Example of a GUI’ section of the page, the Commission provides the example of a ‘Save’ button, but the explanation is ambiguous: it states that information about this function is already available in the language of the country where the device is used, without specifying where this information is provided or which element is deemed not to require translation.
However, it seems reasonable to interpret this as a reference to information already contained in the instructions for use, particularly as the guidance then suggests using an icon whose function is described in those instructions. That example must be assessed in context and does not provide general authorisation to keep the entire interface in English.
There is no consistent approach among national authorities either. In Hungary, the competent authority accepts an English-language interface only if the Hungarian manual explains it in detail with screenshots, and considers it likely that a complex or text-heavy interface will require a Hungarian version.
In Estonia, the authority does not require the entire interface to be translated into Estonian if the instructions for use provide the necessary information, for example through images.
Romania accepts English-language interfaces on devices intended for healthcare professionals only with the user’s written consent, while Slovenia allows software interfaces to be provided in Slovenian or English.
In Poland, devices intended for lay users must have an interface in Polish or use harmonised symbols.
Also bear in mind that, according to MDCG 2018-5 guidance, changing the language of a medical software interface requires a new UDI-DI. That is why it is advisable to decide which languages the interface will include before assigning identifiers, and to plan each new language version as a change to the device.
Conclusion
To answer the question of which languages are required to place medical devices on the market in Europe, you need to consider three factors: where you intend to sell your devices, who will use them and what information you need to provide. Taking these factors together helps you determine which translations are needed and whether any applicable exemptions exist.
At AbroadLink, our life sciences translation services can help you prepare the language versions of your documentation and software. With a clear list of markets, documents and language versions, you can plan the work and maintain translation consistency during device updates.
Josh Gambín is the founder of AbroadLink and leads the company's sales and strategy. With a degree in Biology and a degree in Translation and Interpreting, his background bridges the scientific and linguistic worlds that define our work.
He is a published author in MultiLingual magazine and has participated as conference speaker at leading industry events, including the GALA and tekom conferences, where he shares AbroadLink's perspective on quality, compliance and the responsible use of language technology in regulated sectors.